|
OPERATING INSTRUCTIONS FOR HITACHI H-7000 TEM
(without alignment)
1.
Fill the anticontamination traps with liquid nitrogen
2. Turn the microscope on |
|
(1)
|
Check the right up panel to see if
the microscope is ready. When the microscope is ready all
VACUUM INDICATOR LIGHTS on the right hand desk panel are
green, except for SPEC which should be orange.
(See below)
|
|
|
| |
COL |
GUN |
SPEC |
CAMERA |
| AIR |
|
|
|
|
| CLOSE |
|
|
orange |
|
| EVAC |
green |
green |
|
green |
|
| (2). |
Turn the POWER SWITCH KEY in right lower panel to COL ON. |
|
(3) |
Press READY/OFF button on
left main panel once to light. |
|
(3) |
Select and press desired ACCEL
VOLTAGE button. |
|
(4) |
Press READY/OFF button
again to light. |
| (5) |
Turn FILAMENT control knob on left
main panel gradually and saturate filament. Check current on HV/BEAM meter.
Adjust the current to about 25 ľA with BIAS knob, if necessary. |
|
(6)
|
Turn BRIGHTNESS control knob on
left main panel counterclockwise to focus a spot on screen. Center spot by
manipulating BRIGHTNESS CENTERING knobs on right main panel. Disperse beam
again.
Avoid prolonged exposure of
screen to cross-over beam. |
|
|
|
2.
Specimen observation |
|
|
|
|
|
|
|
|
|
|
|
|
|
Specimen Exchange |
|
1. |
Specimen Loading |
| |
(1). Load the grid onto the grid holder |
| |
(2).
|
Match the specimen holder with
cylinder groove in the specimen stage of the microscope column and insert it until
stops.
|
|
|
Definitely Avoid turning the holder. otherwise air might be
admitted into
the column. |
| |
(3).
|
Turn the specimen chamber
evacuation switch to EVAC. The specimen chamber evacuation lamp
|
|
|
lights up (in yellow). |
| |
(4).
|
After the yellow lamp is lit,
wait until it goes off. Then the SPEC EVAC green lamp of VACUUM
|
|
|
STATE lights up to indicate
completion of preevacuation of the specimen holder.
While the SPECIMEN EVAV green lamp is lit for 15 seconds, the specimen
holder must be inserted.
|
| |
(5). |
Holding the grip of the
specimen holder, turn it clockwise until stops and push it in the axial
direction
until it stops. Further turn it clockwise and insert it until it stops.
|
| |
|
|
2. |
Specimen unloading |
| |
(1). Turn the filament current off. |
| |
(2). Pull the specimen holder
until stops and turn it counterclockwise also until stops. Further pull and
turn
the holder in the same way as above.
|
| |
(3). Turn the specimen chamber evacuation switch
to AIR. After a few seconds, the SPEC AIR red lamp of the VACUUM
STATE is lit to indicate completion of airleak.
|
| |
(4). Pull out the specimen holder straight out.
Pay attention absolutely NOT turn the holder . |
| |
(5). Remove the specimen from the holder. |
| |
|
| 3. Photographing with the Gatan
digital camera |
| |
(1). Spread the beam to cover the smaller
fluorescent screen. |
| |
(2). Depress the TV button (on TV monitor) to
deactivate the film camera. |
| |
(3). Put the cover on |
| |
(4). Lift the lever on the right panel. |
| |
(5). Go to computer using the
GatanMicgraph software. |