Chemistry Department
Department 3838
1000 E. University Avenue
Physical Science Bldg 204
Laramie, WY 82071
Phone: 307-766-4363
Fax: 307-766-2807
Email: chemistry@uwyo.edu
If the sample does not lock, please do the following:
Manual locking:
Make sure that you have the BSMS Control window open. If not, type 'bsmsdisp'. Please go to the LOCK tab.
Click on "Auto LOCK" and wait for about 1 minute. If it locks, fine, if it doesn't, please continue with manual locking.
The goal of locking is to have a lock signal displayed that is symmetrical and centered, not too small or too big, with significant ringing. Probably you have observed this kind of image many times for many successfully locked samples.
If the lock signal shows no signal, or is not centered: BSMS Control Suite/Lock -> Select: Field -> Adjust Field
If the lock signal is too big or small: BSMS Control Suite/Lock -> Select: Lock gain -> Adjust gain. (Optional: Power, but be careful because of power saturation effects)
If the lock signal is not symmetrical: BSMS Control Suite/Lock -> Select: Phase -> Adjust phase.
When you adjust the values within the BSMS Control Suite, please make sure that the cursor does not leave that window. It may be convenient to adjust the stepsize.
If everything is adjusted, push: 'On' within the BSMS Control suite / Lock, and it should lock.
The spectrum may be shifted later. If so, it is possible to calibrated the axis by typing 'cal', which opens a dialog box. Manual calibration is an interactive tool to set the x-axis to the appropriate chemical with known chemical shift (e.g., a solvent).
After performing automated shimming with "topshim"...
...if it says: "not enough valid points', make sure you read the default shim file with 'rsh'.
...if it says: "too many points lost during fit", try: 'topshim convcomp' to compensate for convection currents, especially for non-viscous solvents that are prone to convection currents.
It is possible to do shimming without locking. To do so, read the next section.
While it is easiest to have deuterated solvents as it makes shimming and long-term experiment more stable, it is not essential. Shimming is still recommended and can be done in various ways.
1. Run a 1-D simple proton experiment with the default shim set (type: rsh <enter> and select an appropriate shim file, see also the basic operation instructions) and take a note of the location (in ppm) of a prominent line.
2. gs <enter> runs the experiment in "live" mode, so you can watch the spectrum change while you change the shims at the same time. To do this, open the BSMS window (type bsmdisp <enter>). Turn the LOCK OFF on the BSMS, and press on SHIM.Modify the shims to maximize the proton amplitude and the shape of the signal.
3. Automatic shimming on a proton line.
The command topshim lockoff 1h o1p=x.xx
will turn the lock off, set the proton frequency for x.xx ppm
If this does not work, return to manual shimming on step 2.
These problems manifest themselves quite often and there is no harm in following below instructions.
ii<enter>. If there is an error, repeat.
ii restart <enter>. If there is an error, repeat.
Precautions:
Procedure:
For temperatures between 0 °C and 25 °C (room temperatures), it is best not to use liquid nitrogen as the cooling gas, as the sample heater must be used at high current with the increased risk of heater burn-out and the temperature stability may be poor.
Instead, it is better to let the air that goes to the NMR probe pass through a coil of copper tubing that is immersed in cold solvent such as a water ice, dry ice/solvent or salt/ice cooling bath. Be careful not to spill the contents.
Popular cooling baths include:
Cooling Bath | Temperature in °C |
---|---|
Dry Ice-Benzene | 5 |
Ice Water | 0 |
Ice-Acetone (1:1) | -15 |
Ice-NaCl (1:3) | -20 |
Dry Ice-Acetone | -78 |
Liquid Nitrogen | -196 |
Liquid Helium | -269 |
It is possible to reach other temperatures by using the heater in the edte-window. The heater power should stay low for temperature stability and safety issues (burn-out), so best choose cooling baths close to the desired temperature.
This software is the standard software that is used to acquire data from the Bruker spectrometers and should be familiar to the Bruker NMR users. Bruker has made the Topspin software free for academic users. Please read details at https://www.bruker.com/products/mr/nmr/nmr-software/nmr-software/topspin-faqs.html
The University has a license for this software, details are at: http://www.acornnmr.com/ and at: instrumentation/NMR/60-nmr-basic-operation
This software can process both NMR and MS dfata. Details are at: http://www.mestrelab.com/
Details are at: http://www.acdlabs.com/
Free software for processing of 1D and 2D spectra. Details are at: http://www.umanitoba.ca/chemistry/nmr/spinworks/
Details are at: http://www.inmr.net/
To import Bruker software raw data and processed data, access the following files:
fid (for 1D data)
ser (for 2D data)
1r (for 1D data processed data within the folder pdata/
2r (for 2D data processed data within the folder pdata/
Chemistry Department
Department 3838
1000 E. University Avenue
Physical Science Bldg 204
Laramie, WY 82071
Phone: 307-766-4363
Fax: 307-766-2807
Email: chemistry@uwyo.edu